C100R jolting volumeter, model standardowy

Kategoria:

The C100R Jolting Volumeter, Standard Model, is a single-station rotating and tapping instrument. It determines the bulk density of dried milk by ISO 8967 | IDF 134, Dried milk and dried milk products — Determination of bulk density. It has a 4.3-inch touch display and a built-in receipt printer. Results are expressed to 0.001 g/mL and can be exported as Excel or PDF.

Method and definitions

The method applies to dried whole milk, dried partly skimmed milk and dried skimmed milk, whether instant or non-instant. It also applies to dried whey, dried buttermilk and dried milk-based infant food, and to the same products when milk fat has been replaced by another fat or when the product has been roller-dried instead of spray-dried.

Bulk density is the mass of the test portion divided by its volume under the stated conditions. The unit is grams per millilitre (g/mL).

  • Poured bulk density, ρ₀: mass divided by the volume after the test portion is transferred into the specified cylinder.
  • Bulk density after 100 taps, ρ₁₀₀: mass divided by the volume after 100 taps.
  • Bulk density after 625 taps, ρ₆₂₅: mass divided by the volume after 625 taps. These 625 taps include the first 100 taps.

The calculation is ρ = m / V. m is the mass of the test portion weighed on a balance, and V is the cylinder reading at the corresponding number of taps. The test portion is not weighed on this instrument. The test portion is 100.0 g ± 0.1 g. If 100 g does not fit into the cylinder, use 50.0 g ± 0.1 g.

Before the determination, bring the laboratory sample to 20 °C to 25 °C, mix it, and then weigh it and transfer it to the cylinder. Mix instant dried milk gently so that the particles are not broken.

Specifications

Test station Single
Model C100R
Method ISO 8967 | IDF 134
Tapping action Rotating and tapping
Tap frequency 250±15 taps/min
Drop height 3 mm ± 0.1 mm
Number of taps 100 taps (ρ₁₀₀) and 625 taps (ρ₆₂₅). The 625 taps include the first 100. Tapping stops automatically at the preset count
Measuring cylinder Capacity 250 mL
Test portion 100.0 g ± 0.1 g, or 50.0 g ± 0.1 g if 100 g does not fit
Result g/mL, expressed to 0.001 g/mL
Method repeatability The absolute difference between two single results, obtained by the same operator with the same instrument on identical material in the same laboratory within a short interval, shall not be greater than 0.025ρ g/mL. ρ is the arithmetic mean of the two results
Method reproducibility The absolute difference between two single results, obtained in different laboratories with different operators and different instruments on identical material, shall not be greater than 0.04ρ g/mL. ρ is the arithmetic mean of the two results
Power supply AC 110/220 V, 50/60 Hz

Method repeatability and method reproducibility are the precision of ISO 8967 | IDF 134. They are not the factory specification of an individual instrument.

Check before use

Before a determination, confirm that the drop height and tap frequency are within the tolerances above. Weigh the test portion on a balance with a scale interval of 0.1 g or smaller.

Features

  • 4.3-inch touch display; tapping stops automatically at the preset count;
  • Built-in 58 mm receipt printer;
  • Results expressed to 0.001 g/mL, with Excel and PDF export;
  • Remote software upgrade.

Example of a report

The table shows the report fields and the calculation only. The volumes are assumed values, not a measurement of a milk-powder lot.

Item Example
Method ISO 8967 | IDF 134
Mass of test portion, m 100.0 g
Volume after transfer, V₀ 200 mL
Volume after 100 taps, V₁₀₀ 180 mL
Volume after 625 taps, V₆₂₅ 160 mL
Poured bulk density, ρ₀ = m/V₀ 0.500 g/mL
Bulk density after 100 taps, ρ₁₀₀ = m/V₁₀₀ 0.556 g/mL
Bulk density after 625 taps, ρ₆₂₅ = m/V₆₂₅ 0.625 g/mL

A formal report also identifies the sample and records any operating detail that is not specified in the method and that may have influenced the result.